Bruker-AXS D8 Discover Instrument Configurations
Rules for Reconfiguration
Facility specialist must reconfigure instrument unless user has been trained to do so.
Reservations made for 9:00 AM and 1:00 PM during the week will be reconfigured automatically according to selection made on the reservation calendar, provided the reservation is made before 8:00 AM.
If you require a reconfiguration for a reservation outside of these times, then
1) Make the reservation on the calendar and
2) Contact the specialist to confirm that he will be available for the reconfiguration.
Reconfiguration takes 15 minutes or less.
Instrument Configurations
These represent the most common configurations at this time. Many other configurations are possible and can be arranged.
GADDS - 15 cm
Area detector at medium distance from sample. This allows full range movement of sample cradle with angular coverage of about 35 degrees.
| X-Ray tube position | GADDS1 position |
| Incident beam optics | Göbel Mirror + 1.2mm MES + 0.5mm PS + 0.5mm Snout |
| Sample holder | Vacuum chuck, spring-load, or capillary |
| Diffracted beam optics | none |
| Detector | GADDS area detector |
| Detector Position | 15 cm |
| Variations | none |
GADDS - 10 cm
Area detector at close distance from sample. Sample cradle movement is limited, but angular coverage is about 50 degrees.
| X-Ray tube position | GADDS1 position |
| Incident beam optics | Göbel Mirror + 1.2mm MES + 0.5mm PS + 0.5mm Snout |
| Sample holder | Vacuum chuck, spring-load, or capillary |
| Diffracted beam optics | none |
| Detector | GADDS area detector |
| Detector Position | 10 cm |
| Variations | none |
GADDS - Trans.
Area detector at closest distance from sample. Little or no sample cradle movement possible. Angular coverage is about 60 degrees. A beamstop allows data collection at low values of 2theta.
| X-Ray tube position | Red line |
| Incident beam optics | Göbel Mirror + 1.2mm MES + 0.5mm PS + 0.5mm Snout + beamstop |
| Sample holder | capillary or fiber holder |
| Diffracted beam optics | none |
| Detector | GADDS area detector |
| Detector Position | 7.5 cm |
| Variations | none |
Point Detector XRR
X-Ray Reflectivity configuration. Point detector and auto-absorber allow large dynamic range.
| X-Ray tube position | Red line |
| Incident beam optics | Göbel Mirror + small MES + Auto Absorber |
| Sample holder | Vacuum Chuck |
| Diffracted beam optics | Standard aperture slit assembly (SS, soller slit, DetS) |
| Detector | Point detector with narrow opening slit holder |
| Detector Position | Any |
| Variations | knife edge PS and snout added to reduce beam size |
Point Detector II - beam
Parallel beam XRD (PB - XRD) configuration. Also used for Grazing Incidence XRD (GIXD).
| X-Ray tube position | Just short of 3rd positioning pin |
| Incident beam optics | Göbel Mirror + 0.8 or 1mm MES |
| Sample holder | Vacuum Chuck |
| Diffracted beam optics | 0.35 degree PPC |
| Detector | Detector Point detector with wide opening holder |
| Detector Position | 250 mm |
| Variations | 0.12 degree PPC PS and snout added to reduce beam size |